April 7, 2021
Digital Modular System - Rinse
1-22
Step
Description
10
Displays a
DO NOT OPERATE
tag on the system control panel.
11
Barricade the system surrounding.
12
Remove any media or work units from the system before
commencing work.
1.13.2
Restoring Lockout Tag out
Step
Description
1
Remove all tools from system.
2
Clear and clean the area from any resulting waste.
3
Remove
DO NOT OPERATE
tag from system control panel.
4
Remove lockout and tag out from
Main Disconnect
switch,
pneumatic
supply valve and incoming
DI water
valve.
5
Remove
LOTO
from
Main Disconnect
switch.
Turn the
Main Disconnect
switch to position ‘1’ (
ON
).
6
Turn
ON
the shut-off valve to supply compressed air to system.
7
Turn
ON
or supply incoming (DI) water/chemical supply.
8
Adjust the ball valves according to the label on the valve or refer to
plumbing schematic for the valve status.
9
Inform operator / supervisor about the status of the system.