DMS-R-IM
Installation
April 7, 2021
Digital Modular System - Rinse
3-2
iv.
Insert the settling tool into the anchor and strike with a hammer until the
lip of settling tool meets the top of machine threaded bolt.
v.
Set the bracket over the anchor and insert the properly sized machine
threaded bolt through the bracket and into the anchor.
3.2
System Clearance
It is strongly recommended that the system to have at least
42 inches
of
maintenance work space all around for service accessibility as well as personnel
safety.
Figure 3-1: System isometric front view - space clearance
Left Clearance
42”
Rear Clearance
42”
Front Clearance
42”
Right Clearance
42”