DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-15
Step.
Description
Images illustration
6
To drain water out, return to Manual
Function screen
Press [DRAIN] button (as circled) to
trigger
BVR2
drain pneumatic valve to
turn on.
All wet tanks will begin to drain.
When all tanks are empty, press again
[DRAIN] button to stop drain.
Note
:
Button in green
latched.
Button in aqua/pink
unlatched
Noticed that the indicators are in black
background when they are turned OFF
Note
: Refer to plumbing drawing for
each station drain valve.
Note
: Operator needs to open drain
valve manually.
7
Check and clean tank bottoms.
Remove broken or dropped parts
which may have been left in the tank.
This is necessary to minimize the
possibility of small objects becoming
lodged in a drain line and causing a
service problem.
N/A
8
It is recommended to open supply
valve to flush tank with fresh DI water
for a few minutes.
N/A
9
To start fill, ensure all tanks are clean
and dry. Return to Manual Function
screen and make sure [Drain] button is
unlatched
to
close
drain pneumatic
valve listed below:
BVR2
Note
:
Button in green
latched.
Button in aqua/pink
unlatched
Note
: Operator needs to close drain
valve manually.