Digital Modular System - Rinse
9-8
Step
Description
Images illustration
4
*continue from previous page
In Setup 2 screen, perform setting
for:
Lower Scale For Conductivity
–
Set conductivity lower scale
(mS). Setting range is 0-999.9.
Upper Scale For Conductivity
–
Set conductivity upper scale
(mS). Setting range is 0-999.9.
Offset Setting
– Offset value
from the measured
conductivity (within -10~10) as
compared to the reading
shown on touch screen.
Slope Adjustment
- Gradient
constant (within 0.8-1.2) to be
filled into formula below to
obtain present conductivity
value.
(
Slope Adjustment
X
measured
power output
) +
Offset Setting
.
Note
:
Offset Setting
and
Slope
Adjustment
is
visible to
Admin
only
Press [Next] to go to Setup 3
screen, perform setting for:
Lower Scale For Resistivity –
Set resistivity lower scale
(Mohm).
Upper Scale For Resistivity –
Set resistivity upper scale
(Mohm).