DMS-R-IM
Appendix
April 7, 2021
Digital Modular System - Rinse
9-15
Step
Description
Images illustration
4
*continue from previous page
Press [Next] to go to Setup 5
screen. In Setup 5 screen, perform
setting for:
Lower Scale For Recirculation
Flowrate –
Set recirculation
flowrate lower scale (GPM).
Setting range is 0 – 26.4.
Upper Scale For Recirculation
Flowrate –
Set recirculation
flowrate upper scale (GPM).
Setting range is 0 – 26.4.
Offset Setting
– Offset value
from the measured
recirculation flowrate (within -
5~5) as compared to the
reading shown on touch
screen.
Slope Adjustment
- Gradient
constant (within 0.8-1.2) to be
filled into formula below to
obtain present recirculation
flowrate value.
(
Slope Adjustment
X
measured
power output
) +
Offset Setting
.
Note
:
Offset Setting
and
Slope
Adjustment
is
visible to
Admin
only
*continue to next page