DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-14
systems and subsystems are working perfectly and no leaking or blockage
encountered.
Check all wet stations and perform tanks cleaning, drain and fill water:
Step.
Description
Images illustration
1
Power
ON
the system.
Refer to
section 4.3 Power ON System
for more details.
N/A
2
Ensure ALL stations are in Local mode
(Refer to section
for procedures).
N/A
3
To stop water supply, open
underneath lift up door and close fill
valve listed below:
BVR1 & BVR10
Please refer to plumbing drawing for
supply ball valve number.
Note
: Operator needs to close fill valve
manually.
N/A
4
Press [Manual Function] button at
Main Menu screen to enter Manual
Function screen. Ensure [FILL] button
(as circled) is
unlatched
.
Note
:
Button in green
latched.
Button in aqua/pink
unlatched
Noticed that the indicators are in black
background when they are turned
OFF.
5
Inspect tanks walls for accumulation
of chemical debris or other
contaminants.
If tanks wall requires cleaning, clean
room wipes or the equivalent is
recommended.
N/A