Preventive Maintenance
April 7, 2021
Digital Modular System - Rinse
6-14
6.5.8
Perform PM for Wye-Strainer (Optional Item)
Wye-Strainer is used to catch big particles especially water from factory inlet.
The use of wye-strainer will prevent blockage inside plumbing and avoid
damage to the pump.
Figure 6-11: Wye-strainer and recirculation pump
Step
Description
1
Power OFF the system.
2
Place a container below the Wye-strainer.
3
Close isolation valves immediately.
(Refer to plumbing drawing for more detail)
4
Unscrew the plug slowly to prevent water spillage.
5
Remove the plug slowly.
6
Clean the net (attach to plug) with clean DI water.
7
Fasten the plug back and tighten it.
8
Open all closed isolation valves and switch ON the system.
9
Ensure not water leaking out from Wye-strainer. If so, redo plumbing
work.
6.5.9
General Check
Step
Description
1
Visually inspect all piping connections for leaks. Repair as required.
2
Visually inspect pump seal. Replace if there are signs of leaking.
Wye-strainer
Recirculation
Pump