DMS-R-IM
Preventive Maintenance
April 7, 2021
Digital Modular System - Rinse
6-22
Step
Description
1
Turn off the system and power supply.
2
Remove underneath panel door.
3
Detach the clamp band that connects the main unit to the sample inlet
unit, piping or tank.
4
Clean the prism surface carefully with a soft tissue soaked with warm
water or ethyl alcohol. If the sample solution contains oil or grease,
use ethyl alcohol to ensure the prism surface does not develop a film.
Development of a film on the prism could cause erroneous
measurements.
5
After cleaning is complete, re-attach the CM-800α unit to the sample
inlet unit, piping or tank.
6
Close underneath panel door.
7
Turn on the system and power supply to start cleaning process.
Note
: “Optional item” (Only applicable when this feature is purchased)