Appendix
April 7, 2021
Digital Modular System - Rinse
9-9
Step
Description
Images illustration
4
*continue from previous page
Offset Setting
– Offset value
from the measured resistivity
(measured by other devices) as
compared to the reading
shown on touch screen.
Slope Adjustment
- Gradient
constant (within 0.8-1.2) to be
filled into formula below to
obtain present resistivity
value.
(
Slope Adjustment
X
measured
power output
) +
Offset Setting
.
Note
:
Offset Setting
and
Slope
Adjustment
is
visible to
Admin
only
Press [Next] to go to Setup 4
screen, perform setting for:
Lower Scale For pH –
Set pH
lower scale (pH). Setting range
is 0 - 14.
Upper Scale For pH –
Set pH
upper scale (pH). Setting range
is 0 - 14.
Offset Setting
– Offset value
from the measured pH (within
-3~3) as compared to the
reading shown on touch
screen.
Slope Adjustment
- Gradient
constant (within 0.8-1.2) to be
filled into formula below to
obtain present pH value.
(
Slope Adjustment
X
measured
power output
) +
Offset Setting
.
Note
:
Offset Setting
and
Slope
Adjustment
is
visible to
Admin
only
*continue to next page