DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-23
Step
Description
Images illustration
5
Press on [Misc.]
button on Main
Menu screen. Misc. 1 screen will
prompt.
Set alarm silence timer (min).
Select Local/Remote mode.
Change station date and time if
necessary.
Disable/enable cycle complete
buzzer alert option.
Press [Next] button to go to Misc. 2
screen. In Misc. 2 screen as shown
will prompt.
Configuration:
Set desire ultrasonic generator
power (%) during
Manual
mode
operation (where
[CYCLE
START]
button is not pressed
yet), if necessary.
Set low sonic power alarm set
point (%).
Press to turn ON the respective
station ultrasonic.
Press [Next] button to go to Misc. 3
screen. In Misc. 3 screen,
Set high and low conductivity
limit setting.
Set high conductivity refill
trigger set point and refill cutoff
hysteresis.
Note
: “
*OPTIONAL
” wording will
displayed if it is not purchased by
customer or option is enabled at
Setup 6 screen.
*continue to next page
Misc. 1 screen (DMS-1218 & DMS-1826)
Misc. 2 screen(DMS-1218)
Misc. 2 screen(DMS-1826)
Misc. 3 screen(DMS-1218 & DMS-1826)