DMS-R-IM
Appendix
April 7, 2021
Digital Modular System - Rinse
9-21
9.9
Conductivity Meter Set Point Adjustment
(Optional Item)
NOTE
To change the parameters on this conductivity meter, please
refer to Components Literature & Manual –
Georg Fisher,
Signet 9900 Transmitter literature.
There are
Conductivity Meters
located at station. It displays the real time water
conductivity of the tank. The same reading is displayed in Misc. 3 screen as well
Please refer to the component manuals for details parameter setting procedure.
Figure 9-4: Conductivity Meter
Follow the following steps to adjust conductivity meter set point:
Step
Description
1
Power ON the system.
2
Locate Georg Fisher Conductivity Meter.
3
Press navigation buttons on the panel to configure the meter:
Select Conductivity sensor with the corresponding unit (S for
Conductivity).
Set Conductivity lower scale and upper scale according to
Conductivity meter loop range for 4-20mA.
Conductivity: If lower scale is 0 µS, then set 4mA
SETPNT
to 0; if
upper scale is 2000000 µS, set 20mA
SETPNT
to 2000000.
(Refer to Georg Fisher, Signet 9900 Transmitter literature for detail
operating description)
Navigation keys
Scale
Conductivity
reading
Unit