Installation
April 7, 2021
Digital Modular System - Rinse
3-5
NOTE
Please refer to assembly, plumbing and electrical drawings for
system facilities requirements.
Also, it is strongly recommended to have a copy of those
drawings as reference while performing installation,
operation and maintenance of this system.
Facility connection’s for water supply and unrestricted drain are provided on
utility panels located on the rear of the system. Vent lines for exhausts (if
applicable) should be ducted away from the process area.
The use of unions in the connections is advisable for serviceability. It is
recommended that you install external valves for all
feed lines
connected to the
unit.
An open site type drain is recommended for the gravity waste drain. This is
done to allow free flow and visual confirmation of the discharge. Do not install
an isolation valve in the drain. It is recommended that isolation valves be
installed prior to connection to the utility source.
Item
Requirement
Power capacity
Model DMS-1218: 240V / 3-phase / 60Hz / 16Amps
Model DMS-1826: 240V / 3-phase / 60Hz / 31Amps
Fill
Model: DMS-1218
Port size
: ½” OD S/S
3-4 GPM @ 30-40 PSI
Model: DMS-1826
Port size
: ¾” OD S/S
5-8 GPM @ 30-40 PSI
Drain
Model: DMS-1218
Unrestricted Drain Port size: 1” OD S/S
Model: DMS-1826
Unrestricted Drain Port size: 1-1/2” NPT S/S
Air Supply
Model: DMS-1218
Port size
: ½” OD S/S (*with Auto drain Valve)
1-2 CFM @ 80-90psi
Model: DMS-1826
Port size
: ¾” OD S/S (*with Auto drain valve)
1-2 CFM @ 80-90psi