DMS-R-IM
Appendix
April 7, 2021
Digital Modular System - Rinse
9-5
9.4
Temperature Present Value Offset Setting
Follow the below steps to set temperature PV offset value:
Step
Description
Images illustration
1
Login as Administrator level.
Refer
section 4.5 Login & Logout
for
procedure.
2
Upon login successful, security
level “
15-ADMINIS.”
will be
displayed on Main Menu screen.
3
Press [Temperature Control]
button on Main Menu screen to
go to Temp. Control screen.
4
At the Temp Control screen, set
the temperature offset and slope
adjustment value for each station
accordingly.
Offset Setting
– Offset value for
system measured temperature
compare to actual temperature
(measured by other devices).
Slope Adjustment
- Gradient
constant (within 0.8-1.20) to be
filled into formula below to obtain
present temperature (PV).
Present Temperature (PV) = (
Slope
Adjustment
X
measured
temperature
) +
Offset Setting
.
Note
:
Offset
Setting and
Slope
Adjustment
is
visible to
Admin
only