DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-32
4.10
Recover from Over Temperature
CAUTION
The heater over temperature thermostat should be set higher
than high temperature alarm set point in Temperature Control
screen.
NOTE
When over temperature alarm triggers, process will abort
automatically. Upon the alarm being triggered rectify and
solve the over temperature problem, press
[CYCLE START]
button of the affected station again to start new cycle.
The thermostat is designed to cutoff power supply when the heater is over
heated. Set the thermostat by adjusting the knob. Note that, the thermostat
temperature setting must be higher than the high temperature alarm set point
in the Temperature Control screen.
Follow instructions below to reset the over temperature thermostat.
Step
Description
Images illustration
1
Log into
Engineer
or
Maintenance
level. Refer section
for procedure.
Go to Main Menu screen.
2
On Main Menu screen, press on
[Temperature Control] button to
enter Temperature Control
screen.