Instrument error
Description
Telescope camera
Telescope camera zero point error, relation
between principal point of telescope camera and
crosshair in telescope in Hz and V - option
If the compensator and the horizontal corrections are activated in the instru-
ment settings, every angle measured in the daily work is corrected automati-
cally. Check whether the tilt correction and the horizontal correction are turned
on.
The results are displayed as errors but used with the opposite sign as correc-
tions when applied to measurements.
The following instrument parts can be adjusted mechanically:
•
Circular level on instrument and tribrach
•
Optical plummet - option on tribrach
•
Allen screws on tripod
To get precise measurements in the daily work, it is important:
•
To check and adjust the instrument from time to time.
•
To take high precision measurements during the check and adjust proce-
dures.
•
To measure targets in two faces. Some of the instrument errors are elimi-
nated by averaging the angles from both faces.
During the manufacturing process, the instrument errors are carefully deter-
mined and set to zero. As mentioned above, these errors can change and it is
highly recommended to redetermine them in the following situations:
•
Before the first use
•
Before every high precision survey
•
After rough or long transportation
•
After long working periods
•
After long storage periods
•
If the temperature difference between current environment and the tem-
perature at the last calibration is more than 20°C
Before determining the instrument errors, the instrument has to
be levelled using the electronic level.
The tribrach, the tripod and the underground should be stable
and secure from vibrations or other disturbances.
The instrument should be protected from direct sunlight to
avoid thermal warming.
It is also recommended to avoid strong heat shimmer and air
turbulence. The best conditions are early in the morning or with
overcast sky.
Before starting to work, the instrument has to become acclimatised to the
ambient temperature. Approximately two minutes per °C of temperature differ-
ence from storage to working environment, but at least 15 min, should be
taken into account.
Mechanical Adjust-
ment
Precise Measure-
ments
☞
☞
☞
☞
Settings - TS instrument
263
Содержание Captivate
Страница 1: ...Leica Captivate Technical Reference Manual Version 3 0 English...
Страница 153: ...3 Continue as if creating a new polyline Refer to Create new polyline Several segments Creating Points and Lines 153...
Страница 561: ...Next step Store saves the new coordinate system QuickGrid 561...
Страница 786: ...7 Stake Line To stake out the selected intersection point all delta values must be 0 000 786 Roads Road...
Страница 812: ...Tunnel Profile Editor Profile details view Tunnel Profile Editor Layer details view 812 Roads Tunnel...
Страница 839: ...Field Option Description Offset limit Editable field Maximum horizontal offset from defined pro file Roads Tunnel 839...
Страница 848: ...Field Option Description Scans com pleted Display only Number of scans being measured Number of total scans 848 Scanning...