Troubleshooting
6-47
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0406532
Warning
Load
Product.xml
failed
Product configuration loading
exception. Some configuration
may be missed. Please restart
the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406533
Warning
Get string failed Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0406534
Error
Desktop
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0406535
Error
Desktop
module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0411501
Error
Allocate
memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
Check memory using state
0411502
Error
Allocate shared
memory
segment failed
System shared memory
exception. Please restart the
system. If this issue persists,
please call for service.
There may exist some
unwanted share memory in
the system, or the caller
have no permission.
0411503
Error
Attach shared
memory
segment failed
System shared memory
exception. Please restart the
system. If this issue persists,
please call for service.
Attaches the shared
memory segment identified
by shmid to the address
space of the calling process
failed.
0411504
Error
Init and open a
named
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
The open file numbers
limited by the system or the
process may be reached, or
the caller have no
permission.
0411505
Error
Post a
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Invalid semaphore or the
maximum allowable value
for a semaphore would be
exceeded.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...