Troubleshooting
6-13
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
4. Check the MCB .
0114001
Warning
Camera serial
port UART write
fault
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0114500
Warning
Camera serial
port
transmission
buffer overflow
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0114501
Warning
Camera serial
port RCC time
out
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
Communication between
MCB and camera failure.
1.MCB may be abnormal.
2.Camera may be
abnormal.
3.F7 (fuse) on mother board
(backplane) may be
abnormal.
4.Bundle cable may be
abnormal
5. DC POWER 5V,
+12V AND
+24V(PN:5085801) may be
abnormal.
6. I.I power supply and I.I
may be abnormal.
0114502
Warning
Camera serial
port
unexpected
data error
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
Camera communication
protocol issue
1.Camera may be
abnormal.
0115001
Warning
Flash erase
fault during
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
Flash erase fault during
calibration
1.MCB
0115002
Warning
Flash erase
fault when
saving product
information
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0115003
Warning
Flash write fault
during MA
calibration
Calibration data save
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
1.MCB may be abnormal
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...