Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-16
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0118800
Warning
Exposure is
disabled
because the
room door is
opened
Room door open. Exposure is
disabled. Please close the door
to enable exposure.
N/A
0118801
Warning
Exposure is
automatically
terminated as
the limit of
9'30" for one
shot is reached
Limit of maximum 9'30" for one
shot reached. Exposure is
terminated.
N/A
0118802
Warning
Accumulated
exposure
display time is
longer than
99.9 minutes.
Mainframe
can’t display
rightly
accumulated
exposure time.
Accumulated fluoro time
longer than 99.9 minutes. This
exceeds mainframe fluoro time
display capacity. Please press
and hold Alarm key to reset the
accumulated exposure time.
N/A
0118803
Warning
9'30" exposure
time limit will
be reached in
30"
Limit of maximum 9'30" for one
shot will be reached in 30s.
Exposure will be terminated
then.
N/A
0119500
Warning
Priority II
message buffer
overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
MCU abnormal, MCB may
be abnormal.
STOP0032
768
Warning
WKS error
before MF boot
up
Just display on Mainframe
Control Panel
There is calibration data in
MCB, but WKS didn't
configure product model
Imaging
0205501
Error
Allocate
memory failed
Run out of memory. Please
restart the system. If this issue
persists, please call for service.
Check memory using state
0205502
Error
Allocate shared
memory
segment failed
Memory exception. Please
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted share memory in
the system, or check if
software is installed
correctly.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...