Troubleshooting
6-19
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0205524
Warning
Read IQ param
failed, the value
is over limit
Image processing parameter
over limit. Image quality may
be affected. Please restart the
system. If this issue persists,
please call for service.
check IQ config file
0205525
Warning
Write IQ
parameters
failed
Image processing parameter
save failure. Image quality may
be affected. Please restart the
system. If this issue persists,
please call for service.
check IQ config file
0201501
Warning
Start callback
failure
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
Start capture failed.
1. Gige camera may be
abnormal
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0201502
Warning
End callback
failure
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. Gige camera may be
abnormal
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0201503
Warning
Init camera
buffer failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...