Periodic Maintenance
8-3
X-radiation Exposure
The X-ray tube assembly produces X-radiation when energized. Never operate this device without X-ray
shielding in place. Use lead shielding and draping to protect personnel.
Lead aprons, radiation monitors, and appropriate radiation shields must be utilized for
the protection of all personnel in the vicinity while performing tests. Living human
anatomy should never be used as a phantom or demonstration aid. Do not produce
radiation in hallways.
Many of the PCB's in this system contain components, which are sensitive to Electro-
Static Discharge (ESD). Observe ESD safety procedures.
8.1.2.
Tools and Test Equipment
In addition to standard service tools and test equipment, the following special tools and test equipment
are required to complete the PM procedure.
Standard Tool
Item
Tool
1
Oscilloscope Tektronix 2236 or equivalent
2
Multimeter or equivalent
3
Spatial Resolution Tool
4
Lead Apron
5
kV meter
6
Dosimeter
7
Ampere meter or equivalent
8
Ruler longer than 12 inches (30cm)
9
Erasable marker (whiteboard marker)
10
mAs meter
11
Precision flat head screwdriver (
∅
3mm or less) to adjust camera iris, focus
12
Transparent adhesive tape, with cutter (or need scissor or knife to cut the tape)
Special Tool
Item
Tool
GE part number
1
Three copper sheets, each 1mm thick
00-877682-01
2
9-inch BAFT Tool or UBAT
00-878105-02 or 5307604
3
PWA, extend board
5085609
WARNING
CAUTION
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...