Troubleshooting
6-57
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0407524
Error
MM module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0407525
Error
MM module
asyncNotify
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0407800
Warning
Export images
to CD/DVD
failed
Export to CD/DVD failure. Disk
is write protected or there is no
enough space. Please replace
the disk and try again. If this
issue persists, please call for
service.
Make sure the disc is
writable and there is
enough space.
0407801
Warning
Export images
to USB failed
Export to USB failure. Please
make sure the disk is well
plugged, not write protected,
and has enough space. And
then try again. If this issue
persists, please call for service.
Make sure the USB disk is
well connected and is
writable, do have enough
space.
0407802
Warning
Print image
failed
Print image failure. Please
make sure the printer is ready
for print and try again. If this
issue persists, please call for
service.
Print error: the print job
can’t be finished, please
check the log file for more.
0407803
Warning
Import data
from the disc
failed
Import data from the disk
failure. Please retry or insert
another disk. If this issue
persists, please call for service.
Please insert another disc.
0402502
Warning
Insert exam
failed
Insert exam failure. Please try
again. If this issue persists,
Please call for service.
0402504
Warning
Insert image
failed
Insert image failure. Please try
again. If this issue persists,
Please call for service.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...