Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-54
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0405508
Warning
SCU performs
storage
operation failed
DICOM storage failure. Please
make sure the network and
storage server configuration
are correct, and the server is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405509
Error
SCU sets locale
failed
System start failure. Set locale
failed. Please restart the
system. If this issue persists,
please call for service.
System unknown error
0405510
Error
SCU reports
ready failed
System start failure. Please
restart the system. If this issue
persists, please call for service.
make sure if omniNames
service failed, any module
crashed
0405701
Error
SCU module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
crashed
0405702
Error
SCU module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
crashed
0405703
Error
SCU creates
thread failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
System unknown error
0405704
Warning
SCU cancels
thread failed
Cancel execution thread
exception. Please restart the
system. If this issue persists,
please call for service.
System unknown error
0405705
Error
SCU allocates
memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
Check memory using state
0405706
Error
SCU module
register servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
initialization failed
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...