Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-74
Auto Technique Tracking
Image Resolution
Copper
Filters
(mm)
Specification (kVp)
Measured
(kVp)
Adjustment
done?
Mag Mode
(Field Size)
Specification
(LP/mm)
Measured
(LP/mm)
Adjustment
done?
715/785
(II PN:
5075001)
715/785/
865 (I.I PN:
5085917)
865
(I.I PN:
5075008)
718/785
865
1
64±4
62±4
62±4
Yes No NORM (9”)
≥ 2.0
≥2.2
Yes No
2
74±4
72±4
72±4
Yes No
MAG1 (6”)
≥2.6
≥3.0
Yes No
3
82±4
80±4
80±4
Yes No MAG2 (4.5”)
≥3.1
≥3.5
Yes No
Dose Limitation Check
Dose Group
Normal Fluoro/Low Dose Fluoro
HLF/Low Dose HLF
Dose group: Normal 50 and HLF 50
Max /
≤
47.5 mGy/min
Max /
≤
47.5mGy/min
Dose group: Normal 50 and HLF 100
Max /
≤
47.5 mGy/min
Max /
≤
95 mGy/min
Dose group: Normal 88 and HLF 100
Max /
≤
83.6 mGy/min
Max /
≤
95 mGy/min
Beam Alignment Check
DAP Accuracy Check
Camera centering
Collimator Centering
Dose error <10% DAP error <10% in calibration
Verification
Brivo OEC 715/785 <6 scales
≤4 scales
Dose error <30% DAP error <30% in application
Verification
Brivo OEC 865
≤1
scale
System Status Backup (If needed)
Function Check for Option
Option Description
Verified
Not Installed
Option Description
Verified
Not
Installed
CD/DVD Burner
Video Distributor
I.I. Side Laser Aimer
Tube side Laser Aimer
Printer UP-D897
Printer UP-990-AD
Printer UP-D898
Printer UP-971-AD
Tube Skin Spacer
Film Cassette Holder
Wireless DICOM
Software Options
Pediatric Application:
Removable Filter and Sensor
Removable Grid
Pedal foot switch:
Dual pedal footswitch with save
button
Dual pedal footswitch
Fill PLC Card
Yes No
Reporting
Reporting complete
Yes No
2579 Form Submission ID # # (where applicable)
Calibrated Tools List
Tool Name
Tool Number
Calibration Due Date
Installation Comments
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...