Troubleshooting
6-63
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0401003
Warning
Read serial port
failed
Workstation and mainframe
communication exception.
Please restart the system. If
this issue persists, please call
for service.
0401004
Error
Set serial port
failed
Workstation and mainframe
communication exception.
Please restart the system. If
this issue persists, please call
for service.
0401005
Warning
Communication
Lost
Workstation and mainframe
communication lost. Please
restart the system. If this issue
persists, please call for service.
1. Check AC input power.
2. Check DC power
su5V,+/-15V,24V.
3. Check II power supplier.
4. Check computer power
supplier.
5. Check connections link to
suspected power supplier.
6. Check cables between
computer and PCB-box.
7. Replace MCB.
8. Replace the mother
board (backplane).
0408500
Error
Allocate
memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408501
Warning
Write PIPE
failed
System can’t get peripheral
device information. If this issue
persists, please call for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408502
Warning
Read PIPE failed System can’t get peripheral
device information. If this issue
persists, please call for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...