Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-50
6.
Replace the new Image intensifier on C-Arm.
Be carefully when disassemble the Image
intensifier and make cables pass through
the Image intensifier housing to avoid
Image Intensifier drop.
Be careful not to touch the camera lens or
image intensifier output window and keep
both of them clear without any impurity.
7.
Reassemble Image intensifier power supply, CCD camera, counter weight and connect all cables.
Ensure the cable tightly connect to the Image intensifier Power Supply. False
connection will lead to no image or image vibration in Fluoro.
8.
Refer to 7.3.2 calibration matrix and chapter 4 to do calibration and checks.
9.
After calibration and check, shut down system and reinstall covers removed in above steps.
7.3.15.
Image Intensifier Power supply
There are 2 FRUs for Image intensifier power supply (P/N: 5075009 for 5075929/5075930 and
5085918 for 5075967).
1.
Remove Image intensifier cover, and then disconnect all cables connected to the power supply
and CCD Camera. Remove secure screws of power supply on Image intensifier for
5075929/5075930. Refer to 7.3.14 for the Image Intensifier removal procedure for 5075967.
CAUTION
CAUTION
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...