System Overview
2-29
switch1
connected means
EMERGENCY_SWT has been
pressed
REF_
EMERGENCY_SWT1
The other
termination of
Emergency
switch1
Passive
-
J1-B11 -
EMERGENCY_SWT2 One termination of
Emergency
switch2
Passive
Open/Close J1-C25 EMERGENCY_SWT2 &
REF_EMERGENCY_SWT2 is not
connected means
EMERGENCY_SWT2 has been
pressed
REF_
EMERGENCY_SWT2
The other
termination of
Emergency
switch2
Passive
-
J1-A26 -
Removable Filter
Signals from removable filter to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
PROTECT_A
Removable filter A
sensor signal
Digital
0V/+15V
J1-B31 +15V= Filter removed
PROTECT_B
Removable filter B
sensor signal
Digital
0V/+15V
J1-B30 +15V= Filter removed
Indicator
Signals from MCB to indicator are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
XRAY_INDICATOR_A
Anode of indicator
Passive
-
J1-B8
The indicator will be lighted when a
certain current about 12mA flows
from XRAY_INDICATOR_A through
indicator to XRAY_INDICATOR_B.
XRAY_INDICATOR_B
Cathode of indicator
Passive
-
J1-A8
Program Port
Program port signals are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
SERIAL_TXD
Transmit signal of serial port
Digital
-7V-+7V
J1-A31 -
SERIAL_RXD
Receive signal of serial port
Digital
-7V-+7V
J1-C22 -
Interlock Loop
Interlock loop signals are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
INTERLOCK_LOOP_A One termination of
interlock loop
Passive
-
J2-B16 INTERLOCK_LOOP_A &
INTERLOCK_LOPP_B is connected in
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...