Periodic Maintenance
8-13
8.6.3.
Workstation Checks
Mechanical Movement:
Refer to 3.6.1 for Workstation mechanical movement check.
Adjust the monitor vertical movement annually, according to below procedure:
10N<Force of monitor vertically movement<50N
Monitor vertical movement PM:
1.
For 865 need remove monitor back cover.
2.
Screw the nut to adjust the tightness of the monitor movement in moderate strength.
Brivo OEC 715/785 Back View
Brivo OEC 865 Back View
If friction of monitor horizontal movement is not effective, perform below procedure to
refresh the friction of monitor horizontal movement.
10N<Force of monitor horizontal movement<50N
Monitor horizontal movement PM:
1.
Remove screws on two sides of monitor support
cover.
2.
Remove 2 inside screws.
3.
Rotate the secure nut clock wisely to increase
friction for horizontal rotation
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...