Calibration
4-47
1.
Place the dosimeter probe on the I.I. grid center, and secure it with transparent tape.
2.
Use 2mm copper to cover the beam path on the X-ray tube, and secure it with tape.
3.
Position the C-arm to lateral position.
4.
Make an exposure in Auto Fluoroscopy, and wait until the technique stabilizes. Check if the dose
rate is within:
Brivo OEC 715/Brivo OEC 785(I.I PN: 5075001):
4.1±0.3mR/min (35.92±2.63uGy/min or
2155±157uGy/hr)
Brivo OEC 715/785/865 (I.I PN: 5085917)
:
3.1±0.3mR/min (27.16±2.63uGy/min or
1629±157uGy/hr)
Brivo OEC 865(I.I PN: 5075008):
3.1±0.3mR/min (27.16±2.63uGy/min or 1629±157uGy/hr)
5.
If failed, open the Camera head cover, and carefully adjust the camera iris. Loosen Camera Iris
Securing Screw, turn Camera Iris to smaller index to decrease the stable dose rate, and to bigger
index to increase stable dose rate.
Camera of Brivo OEC 715/785
5075006 Component GigE Camera AM1565
Camera Iris
Adjustment Ring
Camera Iris
Secure Screw
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...