Troubleshooting
6-5
6.3.3.
Message List
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
X-ray Control
0101001
Warning
CPLD firmware
version is
incompatible
with MCB
software
Mainframe hardware version
exception. The mainframe
stopped. Please call for service.
1. MCU or CPLD Firmware
flashed wrong version
2. MCB Issue
0101002
Warning
Camera frame
sync signal is
lost
Camera signal can’t detected.
Please restart the system. If
this issue persists, please call
for service.
1. Camera may be
abnormal.
2. MCB may be abnormal.
3.Maybe Bundle connection
issue or Bundle cable is
abnormal
0101003
Warning
CPLD works
abnormally
after checking
external
watchdog chip
during system
boot up
Mainframe hardware
exception.The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
1. MCB may be abnormal
0102001
Warning
Tube
temperature
sensor's
feedback SPI
line is works
abnormal
Tube temperature monitoring
exception. Exposure is
disabled. Please restart the
system. If this issue persists,
please call for service.
1.MCB may be abnormal
0102002
Warning
Tube
temperature
sensor's
feedback line is
disconnected
Tube temperature monitoring
exception. Exposure is
disabled. Please restart the
system. If this issue persists,
please call for service.
1. Check the thermal switch
on the top of monoblock.
2. Check connections and
bundle cable.
3. Check signal to MCB.
4. Check the monoblock.
0102800
Warning
Tank surface
temperature
has reached
untouch-
temperature
Xray tube housing overheated.
Exposure is disabled. Please do
not touch the housing. Please
wait for the tube to cool down
before making another
exposure.
1. Check the thermal switch
on the top of monoblock.
2. Check connections and
bundle cable.
3. Check signal to MCB.
4. Check the monoblock.
0102803
Warning
Tube is hot
Tube is hot. You may continue
with exposure, but discretion is
advised.
1. Check connections and
bundle cable.
2. Check OIL_TEMP signal
(temperature sensor inside
monoblock) to MCB.
3. Check the monoblock
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...