Troubleshooting
6-45
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0406510
Warning
Simulate key
input failed
Screen saver exception. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406511
Warning
Refresh screen
saver failed
Refresh screen saver failed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406512
Warning
Load setup
configuration
file failed
Setup configuration loading
exception. Some setup may be
missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406513
Warning
Write screen
saver
configuration
file failed
Screen saver configuration
save failed. The screen saver
configuration will be lost.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406514
Warning
Load security
configuration
file failed
Security configuration loading
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406515
Warning
Update security
configuration
file failed
Update security configuration
file failed. Some seeting may
be lost. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406516
Error
Desktop
module register
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406517
Error
Desktop
module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406518
Error
Desktop
module
subscribe sync
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...