Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-54
1.
Power off system.
2.
Remove the rear cover of the C-Arm.
3.
Remove the kV control board according to 7.3.6.
4.
Cut off the binding tie. Disconnect the wires of the
transformer, which are connected to J2.
5.
Loose the screws and replace a new transformer. Connect
the wires and bind the wires with the binding tie to make a
good layout.
6.
Reassemble all the parts.
7.
Power on the system and take exposure. Verify that the
system can work normally.
Main Transformer (P/N: 5085803)
1.
Power off the system and disconnect Workstation’s AC plug from the AC receptacle.
2.
Remove the rear cover, left and right side covers and front cover of the Workstation.
3.
Move out the computer.
4.
Cut off the binding tie. Disconnect the wires of the
transformer, which are connected to J1 and J2.
5.
Loose the secure screws and remove the
transformer.
6.
Replace a new one. Connect the wires and bind
them with binding tie to make a good wire layout.
7.
Power on the system and take exposure. Verify
that the Workstation and C-Arm complete
software boot and initialization without error
reported on their displays. Select the film mode,
take exposure with 80mAs, 110kVp. Verify that the
system can work normally.
Start Up transformer (P/N: 5085804)
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...