Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-64
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0408503
Error
Set PIPE failed
Module init failure. If this issue
persists, please call for service.
0408504
Warning
Read XML file
failed
Setup configuration laoding
failure. Some setup
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
Read setup configuration
file failure
1.Try to import setup
configuration file
2.Reinstall WKS software
3.Replace hardware disk
0408505
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408506
Error
Thread
operation failed
Execution thread exception.
Please restart the system. If
this issue persists, please call
for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408507
Error
Thread sync
operation failed
Exectuion exception. Please
restart the system. If this issue
persists, please call for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408508
Warning
Read default
XML file failed
Default Setup configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
WKS software work
abnormal
1.Reboot system
2.Reinstall system
0408509
Warning
System connect
UPS failed
System can’t detect UPS.
Please shut down system to
avoid data loss. If this issue
persists, please call for service.
Communication between
UPS and WKS is abnormal.
1. PC USB for UPS may be
abnormal
2. Connection between PC
and UPS may be abnormal
3. UPS may be abnormal.
0408510
Warning
UPS return
nothing or
wrong data
System can’t detect UPS.
Please shut down system to
avoid data loss. If this issue
persists. please call for service.
Communication between
UPS and WKS is abnormal.
1. PC USB for UPS may be
abnormal
2.Connection between PC
and UPS may be abnormal
3. UPS may be abnormal.
4. Reinstall WKS software
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...