Calibration
4-45
d.
Connect Camera Cables, and make sure to tighten the secure screws.
e.
Assemble the I.I. cover.
f.
Power on the system.
Ensure the cable tightly connect to the I.I. Power Supply. Virtual connection will lead to
no image or image vibration in Fluoro.
4.6.6.
Limiting Field Size (Collimator Iris Field Size)
Limiting Field Size Verification
1.
Create a new exam.
2.
Load a film cassette in the BAFT.
3.
Make sure the iris is opened completely.
4.
Make an exposure at
in Normal, MAG 1, MAG 2 for 5~10 seconds for each mode.
Take the three exposures on one sheet of film. Note: to select the Film MAG Mode, do the
following: Select
Manual - Fluoro,
then
Normal, MAG1 or MAG 2
.
5.
Remove the Film Cassette and develop the film.
6.
Measure the Normal field size diameter of the collimator iris as projected on the film at 45
°
and
135
°
angle with the ruler.
7.
Set the system to Normal field size, and take an exposure.
8.
Measure, using the BAFT tick marks on the display monitor, the diameter of the circular image at
45
°
and 135
°
angle.
L1
L2
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...