Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-60
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0412503
Error
CORBA fatal
error
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
1. Software internal error
2. Corba server error.
0412504
Warning
Load or parse
setup
configuration
file failed
Setup configuration loading
exception. Some setup
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
1. Configure file miss
2. Configure file format
error
0412505
Warning
Load or parse
administration
configuration
file failed
Administration configuration
loading exception. Please
restart the system. If this issue
persists, please call for service.
1. Configure file miss
2. Configure file format
error
0412506
Warning
Load or parse
version
configuration
file failed
Software version access
failure. Please restart the
system. If this issue persists,
please call for service.
1. Configure file miss
2. Configure file format
error
0412507
Warning
No handler for
the message
Unexpected message. Pleasre
restart the system. If this issue
persists, please call for the
service.
0412510
Warning
Mainframe and
workstation
softawre
incompatible
Mainframe software
incompatible. Mainframe is
stopped. Please make sure
connecting the right
mainframe. If the mainframe is
correct, please call service to
upgrade mainframe and/or
workstation software.
MCB and computer
software version are not
compatible.
1. Check MCB input power
voltage.
2. Reinstall computer
software.
3. Reinstall MCB software.
0412512
Warning
Load or parse
connectivity
setup
configuration
file failed
Connectivity configuration
loading failure. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
0412513
Error
SystemControl
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...