Calibration
4-5
4.2.2.
kV Adjustment
Provide appropriate X-ray protection on surface of Image Intensifier in calibration and
verification to protect Image Intensifier.
1.
Remove the C-arm console cover. Check the connection of IGBT board cables.
2.
Remove the cover of E-Box. Install MCB board with extend board. Remove the J46 jumper on
extend board. Connect the probes of ampere meter to TP222 (+) and TP210 (-) on extend board.
3.
Power on the system. Check that DS9, DS10 and DS3 indicators on kV control board light.
4.
Hold down keys “Ctrl +Shift + Alt +
→
” to invoke the service tool. Login service tool with password.
Click Calibration Tab, then click “mA Calibration” button, “CA11” will display on control panel.
5.
Rotate R40 and R27 on kV control board clockwise to the end. (If applicable)
6.
Test the frequency of TP4 on kV Control Board with Oscilloscope. Adjust R13 on kV Control Board
to make the frequency of TP4 is 80kHz
±
1kHz (TP6 and TP9 are GND.)
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...