Replacement
7-47
Brivo OEC715/785
Brivo OEC865
6.
Install the new monitor and reset all parts and cables. Make sure the binding is still used after the
monitor replacement.
7.
Power on the system. The left monitor of Workstation will display GE logo screen. The right monitor
will enter into the PATIENT INFORMATION Screen automatically. Perform the brightness and contrast
adjustment in section 4.8 if necessary.
7.3.13.
CCD Camera
Dangerous voltage levels are present when covers are removed from the C-Arm.
To prevent ESD damage to the old and new CCD Camera, make sure that you are grounded
with a properly functioning write strap before handling any ESD sensitive electronic
components.
Camera type will be detected during software being installed, and then software will install
corresponding camera driver based on the detection. So from WKS software Ver 1.0,
connect mainframe to WKS to insure the Camera type can be detected during the software
installation process.
1.
Power off the system and position the C-Arm for easy access.
2.
Remove the Image intensifier up cover.
3.
Release the plug secure screws and unplug Camera cables.
4.
Remove the upper counter balance.
5.
Remove 3 camera secure screws and the old camera from the Image intensifier. Pay attention to
original mounting direction.
WARNING
CAUTION
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...