Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-46
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0406519
Error
Desktop
module
unsubscribe
sync event
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406521
Error
Load
application
configuration
file failed
Application configuration
loading exception. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406522
Warning
Load mode
setup
configuration
file failed
Mode Setup configuration
loading exception. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406523
Error
Desktop
module resolve
object
exception
Workstation communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406525
Warning
OSGI
encryption
failed
Account encryption exception.
Account password can’t be
authenticated. Please restart
the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406526
Warning
Dose exceeds
customized
threshold
Dose exceeded the threshold.
0406527
Warning
Load
administration
configuration
file failed
Administration configuration
loading exception. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0406529
Warning
Encounter
undefined
exception
Undefined exception. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0406531
Warning
Restart screen
saver failed
Restart screen saver failed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
Содержание Brivo OEC 715
Страница 2: ......
Страница 19: ...Chapter1 Introduction and Safety...
Страница 41: ...Introduction and Safety 23 46 54 20 18 26...
Страница 55: ...Chapter2 System Overview...
Страница 137: ...Chapter3 Installation...
Страница 147: ...Installation 3 11 7 Put the ramp in position Release all bracket and move the workstation along the ramp off the pallet...
Страница 211: ...Installation 3 75 3 9 Reporting For U S Systems Complete and submit Form 2579 within 15 days of installation...
Страница 212: ...Chapter4 Calibration...
Страница 223: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 12 Mainframe Control Board TP7...
Страница 231: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 20 Mainframe Control Board TP17 POT4 J5 TP7 POT7 POT5...
Страница 275: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 4 64...
Страница 284: ...Chapter5 Software...
Страница 312: ...Software 5 29 4 Click Next 5 Click Install 6 Click Finish button to complete installation procedure...
Страница 326: ...Software 5 43 2 Click on install to continue 3 Click Next to continue...
Страница 335: ...Chapter6 Troubleshooting...
Страница 408: ...Chapter7 Replacement...
Страница 418: ...Replacement 7 11 166...
Страница 488: ...Chapter8 Periodic Maintenance...
Страница 502: ...Periodic Maintenance 8 15...
Страница 505: ...Chapter9 Technical Reference...
Страница 520: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Service Manual 9 16 Lateral configuration 1 5m Lateral configuration 1m...
Страница 521: ...Technical Reference 9 17 Vertical configuration 1 5m Vertical configuration 1m...
Страница 526: ...11 Appendix System Schematics...