Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
85
diagnostic message
S862 Partly filled
. In the case of non-homogeneous media or air
pockets, the damping of the measuring tubes increases.
User entry
• Damping is disabled if
0
is entered (factory setting).
• Damping is enabled if the value entered is greater than
0
.
• The value entered depends on application-specific influence variables, such as the
medium, nominal diameter, sensor etc.
Example
• If the pipe is filled normally the value of the oscillation damping is 500.
• If the pipe is partially filled the value of the oscillation damping is > 5000.
• A practical damping value would then be 2000: enter 2000 as the value.
3.2.4
"Measurement mode" submenu
Navigation
Expert → Sensor → Measurement mode
‣
Measurement mode
Select medium (6062)
Select gas type (6074)
Sound velocity (6147)
Temp. coeff. SV (6181)
Multi-freq.activ
Navigation
Expert → Sensor → Measurement mode → Multi-freq.activ (6242)
Description
Switching the dual mode of the sensor on and off.
Selection
• No
• Yes
Factory setting
No
Select medium
Navigation
Expert → Sensor → Measurement mode → Select medium (6062)
Description
Use this function to select the type of medium.