Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
19
Additional information
Description
If several measured values are displayed at once, the measured value selected here will be
the first value to be displayed. The value is only displayed during normal operation.
The
Format display
parameter (→ 15) is used to specify how many measured
values are displayed simultaneously and how.
Dependency
The unit of the displayed measured value is taken from the
System units
submenu
Selection
•
Oscil. frequency
option
Displays the current oscillation frequency of the measuring tubes. This frequency
depends on the density of the medium.
•
Oscil. amplitude
option
Displays the relative oscillation amplitude of the measuring tubes in relation to the
preset value. This value is 100 % under optimum conditions.
•
Oscil. damping
option
Displays the current oscillation damping. Oscillation damping is an indicator of the
sensor's current need for excitation power.
•
Signal asymmetry
option
Displays the relative difference between the oscillation amplitude at the inlet and outlet
of the sensor. The measured value is the result of production tolerances of the sensor
coils and should remain constant over the life time of a sensor.
0% bargraph 1
Navigation
Expert → System → Display → 0% bargraph 1 (0123)
Prerequisite
A local display is provided.
Description
Use this function to enter the 0% bar graph value to be shown on the display for the
measured value 1.
User entry
Signed floating-point number
Factory setting
Country-specific:
• 0 kg/h
• 0 lb/min
Additional information
Description
The
Format display
parameter (→ 15) is used to specify that the measured value
is to be displayed as a bar graph.
User entry
The unit of the displayed measured value is taken from the
System units
submenu