Proline Cubemass 300 Modbus RS485
Modbus RS485 Register Information
Hauser
283
Navigation: Expert → Diagnostics → Data logging
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Data log. status (0858)
5937
Integer
Read
0 = Done
1 = Stopped
2 = Active
3 = Delay active
Logging duration (0861)
2827 to 2828
Float
Read
Positive floating-point number
*
Visibility depends on order options or device settings
"Simulation" submenu
Navigation: Expert → Diagnostics → Simulation
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Assign proc.var. (1810)
6813
Integer
Read / Write
0 = Off
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow
4 = Density
5 = Ref.density
7 = Temperature
73 = Concentration
74 = Target mass flow
75 = Carrier mass fl.
78 = Carrier vol. fl.
79 = Carr.corr.vol.fl
83 = Oil density
84 = Water density
85 = Density average
86 = GSV flow
87 = GSVa
88 = Oil mass flow
89 = Water mass flow
90 = NSV flow
91 = NSVa
92 = S&W volume flow
93 = Ref.dens.altern.
94 = Oil corr.vol.fl.
95 = Water corr.v.fl.
96 = Targ.corr.vol.fl
97 = Target vol. flow
98 = Temp. average
99 = Oil volume flow
101 = Water cut
102 = Water vol. flow
Proc. var. value (1811)
6814 to 6815
Float
Read / Write
Depends on the process variable selected
Status inp. sim. (1355)
2620
Integer
Read / Write
0 = Off
1 = On
Signal level (1356)
2638
Integer
Read / Write
0 = Low
1 = High
Curr.inp 1 to n sim. (1608–1 to n)
1: 6127
2: 6128
3: 6129
Integer
Read / Write
0 = Off
1 = On
Value curr.inp 1 to n (1609–1 to n)
1: 6139 to 6140
2: 6141 to 6142
3: 6143 to 6144
Float
Read / Write
0 to 22.5 mA
Curr.out. 1 to n sim. (0354–1 to n)
1: 5939
2: 5940
3: 5941
Integer
Read / Write
0 = Off
1 = On