Modbus RS485 Register Information
Proline Cubemass 300 Modbus RS485
270
Hauser
Navigation: Expert → Output → PFS output 1 to n
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Failure mode (0451–1 to n)
1: 2367
2: 2368
3: 4681
Integer
Read / Write
0 = Actual value
1 = 0 Hz
2 = Defined value
Failure freq. (0474–1 to n)
1: 3510 to 3511
2: 3512 to 3513
3: 9908 to 9909
Float
Read / Write
0.0 to 12
500.0 Hz
Output freq. 1 to n (0471–1 to n)
1: 3462 to 3463
2: 3464 to 3465
3: 9910 to 9911
Float
Read
0.0 to 12
500.0 Hz
Switch out funct (0481–1 to n)
1: 3022
2: 3023
3: 9914
Integer
Read / Write
0 = Off
1 = On
2 = Diag. behavior
3 = Fl. direct.check
4 = Limit
5 = Status
Assign diag. beh (0482–1 to n)
1: 3096
2: 3097
3: 9913
Integer
Read / Write
0 = Alarm
1 = Warning
2 = Alarm or warning
Assign limit (0483–1 to n)
1: 3184
2: 3185
3: 4722
Integer
Read / Write
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow
4 = Density
5 = Ref.density
7 = Temperature
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
66 = Pressure
73 = Concentration
74 = Target mass flow
75 = Carrier mass fl.
78 = Carrier vol. fl.
79 = Carr.corr.vol.fl
83 = Oil density
84 = Water density
86 = GSV flow
87 = GSVa
88 = Oil mass flow
89 = Water mass flow
90 = NSV flow
91 = NSVa
92 = S&W volume flow
93 = Ref.dens.altern.
94 = Oil corr.vol.fl.
95 = Water corr.v.fl.
96 = Targ.corr.vol.fl
97 = Target vol. flow
99 = Oil volume flow
100 = Oscil. damping
101 = Water cut
102 = Water vol. flow
Switch-on value (0466–1 to n)
1: 3242 to 3243
2: 3244 to 3245
3: 4728 to 4729
Float
Read / Write
Signed floating-point number
Switch-off value (0464–1 to n)
1: 3234 to 3235
2: 3236 to 3237
3: 4724 to 4725
Float
Read / Write
Signed floating-point number
Assign dir.check (0484–1 to n)
1: 3363
2: 3364
3: 4732
Integer
Read / Write
0 = Off
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow