Proline Cubemass 300 Modbus RS485
Modbus RS485 Register Information
Hauser
235
6
Modbus RS485 Register Information
6.1
Notes
6.1.1
Structure of the register information
The individual parts of a parameter description are described in the following section:
Navigation: navigation path to the parameter
Parameter
Register
Data type
Access type
Selection/input
→
Name of parameter
Indicated in decimal
numerical format
• Float length =
4 byte
• Integer length =
2 byte
• String length,
depending on
parameter
Possible type of access
to parameter:
• Read access via
function codes 03,
04 or 23
• Write access via
function codes 06,
16 or 23
Options
List of the individual
options for the
parameter
• Option 1
•
Option 2
•
Option 3
(+)
• Factory
setting
highlighted in
bold
•
(+)
= Factory
setting
depends on
country, order
options or
device
settings
User entry
Input range for the
parameter
Page number information
and cross-reference to the
standard parameter
description
NOTICE
If non-volatile device parameters are modified via the MODBUS RS485 function codes
06, 16 or 23, the change is saved in the EEPROM of the measuring device.
The number of writes to the EEPROM is technically restricted to a maximum of 1 million.
‣
Make sure to comply with this limit since, if it is exceeded, data loss and measuring
device failure will result.
‣
Avoid constantly writing non-volatile device parameters via the MODBUS RS485.
6.1.2
Address model
The Modbus RS485 register addresses of the measuring device are implemented in
accordance with the "Modbus Applications Protocol Specification V1.1".
In addition, systems are used that work with the register address model "Modicon Modbus
Protocol Reference Guide (PI-MBUS-300 Rev. J)".
Depending on the function code used, a number is added at the start of the register
address with this specification:
• "3" → "Read" access
• "4" → "Write" access