Description of device parameters
Proline Cubemass 300 Modbus RS485
150
Hauser
• Limit
Indicates if a specified limit value has been reached for the process variable. Is used to
output diagnostic information relating to the process and to react to it appropriately at
the system level.
• Fl. direct.check
Indicates the flow direction (forward or reverse flow).
• Digital Output
Indicates the device status depending on whether empty pipe detection or low flow cut
off is selected.
Assign dir.check
Navigation
Expert → Output → Relay output 1 to n → Assign dir.check (0808–1 to n)
Prerequisite
In the
Relay outp.func.
parameter (→ 149), the
Fl. direct.check
option is selected.
Description
Use this function to select a process variable for monitoring the flow direction.
Selection
• Off
• Volume flow
• Mass flow
• Correct.vol.flow
Factory setting
Mass flow
Assign limit
Navigation
Expert → Output → Relay output 1 to n → Assign limit (0807–1 to n)
Prerequisite
In the
Relay outp.func.
parameter (→ 149), the
Limit
option is selected.
Description
Use this function to select a process variable for the limit value function.
Selection
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
• Target vol. flow
• Carrier vol. fl.
• Targ.corr.vol.fl
• Carr.corr.vol.fl
• Density
• Ref.density
• Ref.dens.altern.
• GSV flow
• GSVa
• NSV flow
*
Visibility depends on order options or device settings