Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
99
Additional information
Description
Corrected value = (factor × value) + offset
Temp. offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. offset (1870)
Description
Use this function to enter the zero point shift for the temperature trim. The temperature
unit on which the shift is based is K.
User entry
Signed floating-point number
Factory setting
0 K
Additional information
Description
Corrected value = (factor × value) + offset
Temp. factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. factor (1871)
Description
Use this function to enter a quantity factor for the temperature. In each case, this factor
refers to the temperature in K.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
3.2.8
"Calibration" submenu
Navigation
Expert → Sensor → Calibration
‣
Calibration
Cal. factor (6025)
Zero point (6195)