Description of device parameters
Proline Cubemass 300 Modbus RS485
98
Hauser
Additional information
Description
Corrected value = (factor × value) + offset
Corr. vol factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867)
Description
Use this function to enter a quantity factor (without time) for the corrected volume flow.
This multiplication factor is applied over the corrected volume flow range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
Ref.dens. offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. offset (1868)
Description
Use this parameter to enter the zero point shift for the reference density trim. The
reference density unit on which the shift is based is 1 kg/Nm³.
User entry
Signed floating-point number
Factory setting
0 kg/Nm³
Additional information
Description
Corrected value = (factor × value) + offset
Ref.dens. factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Ref.dens. factor (1869)
Description
Use this function to enter a quantity factor (without time) for the reference density. This
multiplication factor is applied over the reference density range.
User entry
Positive floating-point number
Factory setting
1