Modbus RS485 Register Information
Proline Cubemass 300 Modbus RS485
268
Hauser
Navigation: Expert → Output → PFS output 1 to n
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Assign pulse 1 to n (0460–1 to n)
1: 2461
2: 2462
3: 4685
Integer
Read / Write
0 = Off
1 = Mass flow
2 = Volume flow
3 = Correct.vol.flow
74 = Target mass flow
75 = Carrier mass fl.
78 = Carrier vol. fl.
79 = Carr.corr.vol.fl
86 = GSV flow
87 = GSVa
88 = Oil mass flow
89 = Water mass flow
90 = NSV flow
91 = NSVa
92 = S&W volume flow
94 = Oil corr.vol.fl.
95 = Water corr.v.fl.
96 = Targ.corr.vol.fl
97 = Target vol. flow
99 = Oil volume flow
102 = Water vol. flow
Value per pulse (0455–1 to n)
1: 3034 to 3035
2: 3036 to 3037
3: 4714 to 4715
Float
Read / Write
Signed floating-point number
Pulse width (0452–1 to n)
1: 2836 to 2837
2: 2838 to 2839
3: 4702 to 4703
Float
Read / Write
0.05 to 2
000 ms
Measuring mode (0457–1 to n)
1: 2394
2: 2395
3: 4683
Integer
Read / Write
0 = Forward flow
1 = Reverse flow
2 = Rev. flow comp.
13 = Forward/Reverse
Failure mode (0480–1 to n)
1: 2948
2: 2949
3: 4708
Integer
Read / Write
0 = Actual value
1 = No pulses
Pulse output 1 to n (0456–1 to n)
1: 3082 to 3083
2: 3084 to 3085
3: 4718 to 4719
Float
Read
Positive floating-point number