Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
83
Assign variable
Navigation
Expert → Sensor → Process param. → Partial pipe det → Assign variable (1860)
Description
Use this function to select a process variable to detect empty or partially filled measuring
tubes.
For gas measurement: Deactivate monitoring due to low gas density.
Selection
• Off
• Density
• Ref.density
Factory setting
Off
Low value
Navigation
Expert → Sensor → Process param. → Partial pipe det → Low value (1861)
Prerequisite
One of the following options is selected in the
Assign variable
• Density
• Ref.density
Description
Use this function to enter a lower limit value to enable detection of empty or partially filled
measuring tubes. If the measured density falls below this value, monitoring is enabled.
User entry
Signed floating-point number
Factory setting
200
Additional information
User entry
The lower limit value must be less than the upper limit value defined in the
High value
parameter (→ 83).
The unit depends on the process variable selected in the
Assign variable
parameter
Limit value
If the displayed value is outside the limit value, the measuring device displays the
diagnostic message
S862 Partly filled
.
High value
Navigation
Expert → Sensor → Process param. → Partial pipe det → High value (1858)
Prerequisite
One of the following options is selected in the
Assign variable
• Density
• Ref.density