Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
129
Value per pulse
Navigation
Expert → Output → PFS output 1 to n → Value per pulse (0455–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected, and one of
the following options is selected in the
Assign pulse
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
Description
Use this function to enter the value for the measured value that a pulse is equivalent to.
User entry
Signed floating-point number
Factory setting
Depends on country and nominal diameter→ 229
Additional information
Entry
Weighting of the pulse output with a quantity.
The lower the pulse value, the
• better the resolution.
• the higher the frequency of the pulse response.
Pulse width
Navigation
Expert → Output → PFS output 1 to n → Pulse width (0452–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected, and one of
the following options is selected in the
Assign pulse
• Mass flow
• Volume flow
• Correct.vol.flow
• Target mass flow
• Carrier mass fl.
Description
Use this function to enter the duration of the output pulse.
User entry
0.05 to 2
000 ms
Factory setting
100 ms
Additional information
Description
• Define how long a pulse is (duration).
• The maximum pulse rate is defined by f
max
= 1 / (2 × pulse width).
• The interval between two pulses lasts at least as long as the set pulse width.
• The maximum flow is defined by Q
max
= f
max
× pulse value.
• If the flow exceeds these limit values, the measuring device displays the diagnostic
message
S443 Pulse output 1 to n
.
*
Visibility depends on order options or device settings