Proline Cubemass 300 Modbus RS485
Modbus RS485 Register Information
Hauser
263
Navigation: Expert → Sensor → External comp.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Temperature mode (6184)
5515
Integer
Read / Write
0 = Internal value
1 = External value
11 = Current input 1
12 = Current input 2
External temp. (6080)
2507 to 2508
Float
Read / Write
–273.15 to 99
999 °C
*
Visibility depends on order options or device settings
"Calculated value" submenu
"Corr. vol.flow." submenu
Navigation: Expert → Sensor → Calculated value → Corr. vol.flow.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Corr. vol.flow. (1812)
5129
Integer
Read / Write
0 = Calc ref density
1 = Fix ref.density
11 = Current input 1
12 = Current input 2
Ext. ref.density (6198)
2509 to 2510
Float
Read / Write
Floating point number with sign
Fix ref.density (1814)
5130 to 5131
Float
Read / Write
Positive floating-point number
Ref. temperature (1816)
5136 to 5137
Float
Read / Write
–273.15 to 99
999 °C
Linear exp coeff (1817)
5132 to 5133
Float
Read / Write
Signed floating-point number
Square exp coeff (1818)
5134 to 5135
Float
Read / Write
Signed floating-point number
*
Visibility depends on order options or device settings
"Sensor adjustm." submenu
Navigation: Expert → Sensor → Sensor adjustm.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Install. direct. (1809)
5501
Integer
Read / Write
0 = In arrow direct.
1 = Against arrow
"Zero point adj." wizard
Navigation: Expert → Sensor → Sensor adjustm. → Zero point adj.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Zero point adj. (6196)
5121
Integer
Read / Write
0 = Cancel
1 = Start
2 = Zero adjust fail
8 = Busy
Progress (2808)
6797
Integer
Read
0 to 100 %
"Variable adjust" submenu
Navigation: Expert → Sensor → Sensor adjustm. → Variable adjust
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Mass flow offset (1831)
5521 to 5522
Float
Read / Write
Signed floating-point number
Mass flow factor (1832)
5519 to 5520
Float
Read / Write
Positive floating-point number