Description of device parameters
Proline Cubemass 300 Modbus RS485
56
Hauser
Output curr. 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n → Output
curr. 1 to n (0361–1 to n)
Description
Displays the current value currently calculated for the current output.
User interface
0 to 22.5 mA
Measur. curr. 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → Value curr.out 1 to n
→ Measur. curr. 1 to n (0366–1 to n)
Description
Use this function to display the actual measured value of the output current.
User interface
0 to 30 mA
"Pulse/frequency/switch output 1 to n" submenu
Navigation
Expert → Sensor → Measured val. → Output values → PFS output
1 to n
‣
PFS output 1 to n
Output freq. 1 to n (0471–1 to n)
Pulse output 1 to n (0456–1 to n)
Switch status 1 to n (0461–1 to n)
Output freq. 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Output
freq. 1 to n (0471–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected.
Description
Displays the actual value of the output frequency which is currently measured.
User interface
0.0 to 12
500.0 Hz